  • TG-142: Complete LINAC QA Guide (AAPM Protocol)

    📋 AAPM TASK GROUP 142

    TG-142: Complete LINAC QA Guide (AAPM Protocol)

    The international reference protocol for radiotherapy quality assurance — tests, tolerances, frequencies and recommended equipment

    2009
    Published by the AAPM
    3
    Test frequencies: daily, monthly, annual
    3
    Tolerance levels: non-IMRT, IMRT, SRS/SBRT
    100+
    Individual tests recommended

    What is TG-142?

    The AAPM Task Group 142 (TG-142) is the reference report from the American Association of Physicists in Medicine that defines the quality assurance (QA) programme for medical linear accelerators (LINACs). Published in 2009, it replaced and significantly expanded the TG-40, becoming the international standard for radiotherapy QA.

    • Updated TG-40 (1994) recommendations
    • Added specific tests for multi-leaf collimators (MLC)
    • Included QA for IMRT, VMAT, and stereotactic techniques (SRS/SBRT)
    • Defined tolerances for integrated imaging systems (CBCT, EPID, kV)
    • Established test frequencies: daily, monthly, and annual

    Why is TG-142 Important?

    Compliance with TG-142 ensures radiotherapy treatments are delivered with maximum precision and safety. A robust QA programme based on TG-142:

    • Ensures dosimetric accuracy — essential for tumour control
    • Detects deviations before they affect patients
    • Meets regulatory and hospital accreditation requirements
    • Supports advanced techniques like IMRT, VMAT and SRS/SBRT safely
    • Meets requirements of the European Directive 2013/59/Euratom on radiation protection

    Daily Daily Tests

    Daily tests verify basic accelerator functionality before starting treatments. With equipment like the Daily QA 4 Pro, these checks can be completed in under 5 minutes.

    ParameterNon-IMRT ToleranceIMRT ToleranceSRS/SBRT Tolerance
    Output (Dose Rate)±3%±3%±3%
    Field Size±2 mm±2 mm±2 mm
    Alignment Lasers±2 mm±2 mm±1 mm
    Distance Indicator (ODI)±2 mm±2 mm±2 mm
    Safety: door, audio/videoFunctionalFunctionalFunctional
    💡 Recommended equipment: The Daily QA 4 Pro from Sun Nuclear checks output, symmetry, energy, field size and uniformity in a single setup.

    Monthly Monthly Tests

    Monthly tests evaluate more detailed beam and mechanical system parameters. The IC Profiler and MapCHECK 3 are reference equipment for these checks.

    ParameterNon-IMRT ToleranceIMRT ToleranceSRS/SBRT Tolerance
    Output Constancy (photons)±2%±2%±2%
    Output Constancy (electrons)±2%±2%N/A
    Profile Symmetry±3%±2%±2%
    Flatness±2%±2%±1%
    Beam Quality (energy)±2%±2%±1%
    Mechanical Isocentre±2 mm diameter±2 mm diameter±1 mm diameter
    MLC Accuracy±2 mm±1 mm±1 mm
    CBCT Image QualityBaselineBaselineBaseline

    Annual Annual Tests

    Annual tests are the most comprehensive and include absolute calibration per the TG-51 protocol. Equipment such as the PC Electrometer and SunScan 3D are essential.

    ParameterGeneral ToleranceSRS/SBRT Tolerance
    Absolute Calibration (TG-51)±1%±1%
    PDD/TMR Verification±2%±2%
    Field Factors±2%±2%
    Light-Radiation Coincidence±2 mm±1 mm
    Mechanical-Radiation Isocentre±2 mm±1 mm
    Collimator Accuracy±2 mm±1 mm
    MLC TransmissionBaseline ±0.5%Baseline ±0.5%
    MLC Leaf SpeedPer manufacturer specifications

    MLC and IMRT-Specific Tests

    TG-142 includes dedicated sections for MLC quality assurance, essential for modern techniques such as IMRT and VMAT.

    MLC/IMRT TestFrequencyIMRT Tolerance
    Leaf Positional AccuracyMonthly±1 mm
    Leaf SpeedMonthlyPer manufacturer spec
    Inter-leaf TransmissionAnnual≤2% of reference
    Tongue-and-groove EffectAnnualPeriodic verification
    Picket Fence TestMonthly±1 mm
    VMAT VerificationMonthlyBaseline

    Imaging Systems QA (IGRT)

    A major innovation of TG-142 over TG-40 is the inclusion of QA recommendations for integrated imaging systems:

    • CBCT (Cone-Beam CT): Image quality, geometric accuracy, patient dose
    • kV Imaging: Spatial calibration, image quality, isocentre coincidence
    • EPID (Electronic Portal Imaging): Uniformity, spatial resolution, calibration
    • Image-isocentre coincidence: ±2 mm (±1 mm for SRS)

    Action Levels in TG-142

    🔍

    Level 1 — Inspection

    Document and monitor trends. Repeat measurement to confirm.

    🔧

    Level 2 — Scheduled Action

    Correct within a defined timeframe. Assess impact on ongoing treatments.

    🛑

    Level 3 — Immediate Action

    Suspend treatments until correction and verification. Do not treat patients.

    Recommended Equipment for TG-142 Compliance

    Sun Nuclear, represented by Interphysix in Europe, offers comprehensive solutions for implementing a TG-142-compliant QA programme:

    Daily QA

    ☀️ Daily QA 4 Pro

    Checks output, symmetry, energy, field size and uniformity in a single setup — results in under 5 minutes.

    View product →
    Monthly QA

    📊 IC Profiler

    2D ionisation chamber array for beam profile, symmetry and flatness measurements.

    View product →
    Monthly QA

    🎯 MapCHECK 3

    High-resolution 2D diode array for IMRT plan QA with automatic gamma analysis.

    View product →
    3D Verification

    🌀 ArcCHECK

    Cylindrical 3D phantom with 1386 detectors for VMAT and IMRT verification.

    View product →
    Dosimetry

    ⚡ PC Electrometer

    High-precision electrometer for absolute calibration per TG-51.

    View product →
    Software

    💻 OneQA

    Centralised platform for QA programme management and automation.

    View product →

    TG-142 vs TG-40: Key Differences

    ❌ TG-40 (1994)

    • Conventional radiotherapy only
    • Basic MLC mention
    • No imaging systems QA
    • Uniform tolerances
    • No SRS/SBRT coverage

    ✅ TG-142 (2009)

    • IMRT, VMAT, SRS, SBRT included
    • Dedicated MLC test section
    • CBCT, kV, EPID covered
    • Technique-specific tolerances
    • Tighter SRS/SBRT tolerances