Daily linac QA in one indexed device
Daily QA™ 4 Pro from Sun Nuclear simplifies daily machine QA by combining dosimetry, Winston-Lutz and imaging/positioning checks into a single indexed, imageable device. The approach reduces phantom swaps, room entries and setup variability.
What it consolidates in morning QA
- Dose constancy, field size, flatness, symmetry, beam shape, penumbra and dose per pulse.
- Flatness and symmetry fields from 2×2 cm to 20×20 cm, and dose constancy from 5×5 cm to 20×20 cm.
- Integrated active Winston-Lutz measurement, without switching to a dedicated phantom.
- Imaging and positioning checks, including CBCT, kV/kV, MV/MV and SGRT verification support.
SunCHECK v6.0+ integration
Automated data capture, consistent documentation and centralised reporting require SunCHECK v6.0 or higher. The configuration can be designed for local/departmental use or for an enterprise architecture, depending on the hospital’s infrastructure and QA workflow.
This distinction matters in proposals: the device performs the integrated measurements, but the automation level depends on the installed SunCHECK version and architecture.
Key technical specifications
| Detectors | 4 vented ion chambers, 225 unattenuated SunPoint® 2 diodes and 24 attenuated SunPoint® 2 diodes |
| Imaging targets | 8 tungsten carbide spheres, 3 mm each |
| Photon energies | 4 MV to 25 MV, FFF and Cobalt-60, according to Sun Nuclear specifications |
| Electron energies | 4 MeV to 25 MeV |
| Accessories | Wireless BaseStation, 6DOF plate, GMF Mounting Fixture, GMF Adapter Plate and user-replaceable battery |
Public resources
See the public Sun Nuclear overview, the Daily QA 4 Pro datasheet and the SunCHECK release highlights.
In Portugal: Interphysix represents Sun Nuclear and supports demonstration, configuration, training and integration into the department’s QA programme. Contact us for a quotation or demonstration.