Easily test all critical functions with the QA-ES III electrosurgery analyzer and tester. Quickly collect all measurements, including vessel sealing, contact quality monitoring (CQM), high frequency (HF) leakage, and output power distribution (single or continuous mode) with the easy-to-use interface guide.
Main Features:
Test all critical Electrosurgical Unit (ESU) functions with accurate ranges of power, current, frequency, crest factor, and load resistance
Collect measurements in single and continuous mode
Meet all global standards including ANSI/AAMI and IEC**
All-in-one: All the hardware and software tools needed to complete preventative maintenance and troubleshooting are built into the unit, eliminating the need to purchase or carry additional cables, wires, switch boxes, and RECM boxes
Easy-to-use interface: Large buttons and LCD screen guide the user through test sequences
Memory storage of up to 5,000 test records, eliminating the need to download data after each preventive maintenance or troubleshooting session is completed
Bipolar leakage test performed with a fixed load of 200 ohms.
Test all critical functions of the Electrosurgical Unit (ESU) with accurate ranges of power, current, frequency, crest factor, and load resistance, ensuring ESU safety. Ensure the performance and safety of electrosurgical units with simplified testing using the QA ES III electrosurgery analyzer and tester. The QA-ES III is capable of testing all modern high-power electrosurgical units, with generator output current accuracy as low as ±2.5%. Collect all measurements, including vessel sealing, contact quality monitor (CQM), high frequency (HF) leakage, and output power distribution in single or continuous mode. You don't need to carry additional accessories or cables, as the QA-ES III has all the hardware and software features necessary for comprehensive testing. With all-in-one capabilities, the QA-ES III is one of the most user-friendly electrosurgery analyzers on the market today. Additionally, optional Ansur test automation software allows users to automatically create and run tests, capture data, and produce easy-to-read reports.
10 Best Practices for Electrosurgical Unit Testing Continuous operation: Continuous measurement of power, current, peak-to-peak voltage (closed load only), and crest factor Single operation: Single measurement after the defined delay time of the ESU output of power, current, peak-to-peak voltage (closed load only), and crest factor Power distribution: Tests the impedance sensing circuitry in power assurance functions in new-generation ESU units. Applicable parameters—power, current, peak-to-peak voltage (closed load only), and crest factor—can be easily observed during measurements RF leakage current: Provides connections and load configurations for measuring HF leakage from grounded and isolated equipment RECQM: Tests the quality monitoring of the return electrode control using the QA-ES internal loads.